Conductivity Cell TB27

Conductivity Cell TB27

The Conductivity Cell TB27 can be inserted or removed from process lines or vessels via a ball valve without disturbing the process. The TB27 sensor is designed for applications that exceed standard hot-tap sensor pressure ratings and for operator safety. An extraction housing isolates the operator from the process fluid. This housing has 1/4 in. ports for flushing, draining, pressurizing, or depressurizing this chamber. Ruggedly constructed of 316 stainless steel, these sensors withstand the most demanding processes and measurement requirements.
  • High pressure hot-tap sensor
  • Measurement ranges from 0 ... 1.999 µS/cm to 0 ... 19.99 mS/cm
  • Corrosion resistant 316 stainless steel measurement electrodes
  • Pt100 or Pt1000 integral temperature compensation elements
  • Maximum temperature 200 °C (392 °F)
  • Maximum pressure 2,068 kPa (20,68 bar / 300 psi)
  • Interchangeable and replaceable sensor tips